RadiCS
Chapter 7 History
56
English
History List
Date & Time Date and time when a task is performed appear.
Task Types of tasks appear.
CAL (Optical sensor) ...........calibration with an optical sensor
CAL (Backlight sensor) .......check with a backlight sensor
CHK (Pattern) .....................pattern check
CHK (Luminance) ...............luminance check
CHK (Grayscale) .................grayscale check
CHK (Uniformity) .................uniformity check
ToneCurve Tuning ...............ToneCurve Tuning
Diagnostics ........................diagnostics
LUT-CAL (Optical sensor) ...to write in results of CAL (Optical sensor) into
LUT
Items with exclamation points “!” will be identified as important and
will stay in the list when reducing history.
Information “Multi-monitor” is displayed when multiple monitors check has been
executed.
Judgment Results of judgment appear.
Passed .......... when a result is accepted.
Failed ............ when a result is rejected.
Canceled ....... when a task is canceled.
Error .............. when an error occurs.
A question mark “?” is added to the category for judgment if
comments are added to the judgment when performing a task.
“-” is added to a record no standard applies.
Standard Standards that were used for a task performance appear.
AAPM-P(A) ........ Acceptance test AAPM Primary standard
AAPM-S(A) ........ Acceptance test AAPM Secondary standard
AAPM-P(C) ........ Constancy test AAPM Primary standard
AAPM-S(C) ........ Constancy test AAPM Secondary standard
Basic QC(A) ....... Acceptance test Basic QC standard
Basic QC(C) ....... Constancy test Basic QC standard
DIN-A(A) ............ Acceptance test DIN Application Category A standard
DIN-B(A) ............ Acceptance test DIN Application Category B standard
DIN-A(C) ............ Constancy DIN Application Category A standard
DIN-B(C) ............ Constancy DIN Application Category B standard
DMGQCM-M(A)
..Acceptance Test DMG QC Manual
standard
DMGQCM-M(C)
..Constancy Test DMG QC Manual
standard
EUREF-MP(A) ... Acceptance test EUREF Primary standard
EUREF-MS(A) ... Acceptance test EUREF Secondary standard
EUREF-MP(C) ... Constancy test EUREF Primary standard
EUREF-MS(C) ... Constancy test EUREF Secondary standard
IEC(C) ................ Constancy IEC standard
JIS(C) ................. Constancy JIS standard
IPEM(A) ............. Acceptance test IPEM standard
IPEM(C) ............. Constancy test IPEM standard
JESRA-1(A) ....... Acceptance test JESRA Grade 1 standard
JESRA-2(A) ....... Acceptance test JESRA Grade 2 standard
JESRA-1(C) ....... Constancy test JESRA Grade 1 standard
JESRA-2(C) ....... Constancy test JESRA Grade 2 standard
QS-RL-A(A) ........ Acceptance test QS-RL Application Category A standard
QS-RL-B(A) ........ Acceptance test QS-RL Application Category B standard
QS-RL-A(C) ....... Constancy test QS-RL Application Category A standard
QS-RL-B(C) ....... Constancy test QS-RL Application Category B standard
QS-RL-AM(A) ..... Acceptance test QS-RL Application Category A
Mammo(PAS1054) standard
QS-RL-AM(C) .... Constancy test QS-RL Application Category A
Mammo(PAS1054) standard
ACR-M(A)........... ACR Mammo acceptance test standard
ACR-M(C) .......... ACR Mammo constancy test standard
CUSTOM(A) ....... Customized acceptance test standard
CUSTOM(C) ....... Customized constancy test standard
Tester The entered tester’s name appears when performing a task.
Use Applications registered in the monitor information appear.
If “//” and letter strings are typed in, letter strings after “//” will not
appear when printing.
Monitor The name of a manufacturer registered in the monitor information
appears with “Model Name S/N: Serial Number” format.