RadiCS
Chapter 4 Acceptance Test / Constancy Test
35
English
4-2. Constancy Test
Monitors for diagnostic imaging use must be tested regularly upon display performance.
(= constancy test) A constancy test has pattern/luminance/gray scale/uniformity. The check
items and judgment standards vary depending on a constancy standard.
The constancy test includes eleven different standards: DIN, AAPM, ACR, IEC, JIS,
JESRA, QS-RL, EUREF, IPEM, Basic QC and DMG QC Manual. Test items can be
customized in accordance with requirements.
To perform a constancy test, the task setup wizard must be on and a constancy test
standard must be set in advance.
To change settings, start up task setup wizard by clicking [Configuration] - [Wizard] in the
advanced mode.
How to Customize the Constancy Test Setting
Note
This operation can only be carried out in the advanced mode.
[Procedure]
1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting
dialog appears. Click [Monitor List] to display the monitors connected to the
computer. Click the right mouse button on the monitor and select “Constancy
test settings”.
Select a sensor.
* When customizing,
Patten/ Luminance/
Grayscale/Uniformity