
RadiCS
Chapter 3 Monitor Calibration
21
English
Display Function
DICOM Part14
GSDF
The setup complies with DICOM Part14.
The monitor luminance is set up being reflected from the
environment light when calibrating. It will also be reflected from
environment light when checking luminance and grayscale.
Reference:
Lmax+Lamb = Maximum Luminance Target
Lmin+Lamb = Minimum Luminance Target
Enter from 0.00 to 9.99 to set up the monitor luminance.
CIE Use display functions in compliance with CIE LUV and CIE LAB.
Exp Enter from 1.6 to 2.4 in the gamma value.
Log Linear Use log linear functions.
Linear Use linear functions.
User Definable The file can be selected by clicking “Add to the list...”.
Tips
• When clicking [Measure], the Color Measurement window appears.
Select a monitor and measure the color.
6. Click [OK] after setup to close the calibration setting window and then click [OK]
on the monitor.
Comentários a estes Manuais