RadiCS
Chapter 4 Acceptance Test / Constancy Test
38
English
5. Open [Grayscale] and set up the error check.
Execute Error
Check
The check box must be valid when grayscale is scanned.
Target Error Rate
(%)
Enter 0 to 100 in the error rate when the ratio to the screen
measurement value (cd/m²) is calculated.
Enter 3 to 256 in the screen measurement point. Enter 0-100
in the error rate when the ratio to GSDF is calculated.
- JND s /Luminance Interval
Measure 256 points, and evaluate the JND per grayscale
difference. Enter the judgment value for each item from 0.0
to 3.0.
Note
The number of measuring points for the grayscale check using RadiCS SelfQC is fixed to
18 points.
C
hanging the measuring points and JNDs/Luminance interval settings for RadiCS
SelfQC is disabled.
Tips
When the grayscale check is executed for RadiCS SelfQC, the history will be saved as
Custom(c) regardless of the selected standard.
Use RadiCS for the final judgment and report generation of the selected standard.
6. Open [Uniformity] to set up the measurement standard.
Number of
Measurement
Point
Select a measurement point of the screen.
Window Size (%) Set up the measurement window size for 5-50.
Judgement
(Luminance
Uniformity)
Set up the judgement standard of luminance uniformity. An error
judgement standard can be set for 2 grayscale uniformities.
The check box must be valid when it is scanned.
Judgement
(Color Uniformity)
Set up the judgement standard of white uniformity.
7. Click [OK] to close the constancy test window and click [OK] on the monitor
when all the tab setups are completed.
Sensor Setting
Procedure
1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting
dialog appears. Click “Sensor” and specify a sensor for constancy test.
2.
To use a EIZO sensor, set up monitor luminance.
See 5-1. Ambient Luminance Input”.
Note
Set up LS-100, CD-Lux, CD mon, LX-Plus or LX-Can as a sensor if selecting DIN or
QS-RL for the acceptance test standard. An acceptance test cannot be performed with
an EIZO sensor.