RadiCS
Chapter 2 RadiCS Setup
17
English
Monitor Information Window
Register this monitor Disable the check box when not controlling the monitor.
Monitor Type Select the monitor type from Color LCD, Monochrome LCD,
Color CRT, or Monochrome CRT.
Select Monitor Displays the “Select Monitor” dialogue. When you select a
monitor from the list, its information is automatically entered.
Manufacturer/
Model Name/
Serial Number (S/N)
Enter the manufacturer, model name, and serial number (S/N).
Asset Number Enter the asset management number.
Installed on Select the date of introduction.
CAL Switch
Mode1 to Mode3
If only a single mode can be calibrated, entries are only
available for Mode1. If several modes can be calibrated,
entries are available for Mode1 to Mode3. The number of
modes depends on the model of the RadiForce monitor.
To manage a mode with RadiCS LE software, enable the
“Control” check box.
Note
Only Mode1 can be set when GS521-ST is used.
Use Select from CT, DR, DSA, MMG, MR, NM, PACS or US. You
can enter any strings in the “Use (Comment)” field.
Back Returns to the monitor type judgment dialogue.
Next Goes to the monitor setup wizard of the next monitor when
setting Wide View or Mirroring.
4. When using a Clip-On Swing Sensor, click [Next >] and make the settings.
When not using a Clip-On Swing Sensor, proceed to step 9.
5. Install the Clip-On Swing Sensor by aligning it to the center of the displayed
pattern. Click [Proceed] to have the software detect the monitor with the Clip-
On Swing Sensor.