RadiCS
Chapter 3 Monitor Calibration
25
English
How to Calibrate a Monitor
Tips
Refer to page 27 when calibrating an EIZO FlexScan M series monitor or a monitor made by
another manufacturer.
[Procedure]
1. Turn the monitor on 30 minutes before calibration.
2. Attach an EIZO sensor (UX1).
3. Click [Proceed] in calibration.
The monitor can also be calibrated by clicking [Calibration] - [Optical sensor] on the
menu bar.
4. [Input tester name] will appear. Input tester’s name and then click [OK].
Note
Input tester’s name within 31 characters.
5. A calibration message and a measurement window are displayed on the
monitor screen.