RadiCS
Chapter 5 Monitor Setup
42
English
Chapter 5 Monitor Setup
The advanced mode has some setups and functions for RadiCS software besides acceptance test
setting and calibration setting. Each connected monitor can be set on its own.
[Procedure]
1. W
hen selecting [Configuration](Advanced mode)/[Configuration/View](User
mode) on the menu bar and then clicking [Change Settings], the setting dialog
appears. Click [Monitor List] to display the monitor list screen.
The connected monitors in the list are displayed of graphics boards.
Tips
•
If more than one mode can be calibrated for the connected monitor, the monitor is
displayed for each mode.
The mode which can be calibrated is identified with the model name (mode number).
2. Select a monitor or the monitor for a specific mode from the list and click the
right mouse button.
Functions
Control this
monitor*
Decide to have a monitor managed by RadiCS or not.
Calibration
settings*
Set up for calibration
See 3-1.
Acceptance test
settings*
Set up for acceptance test.
See 4-1. “How to Customize the Acceptance Test Setting”.
Constancy test
settings*
Set up for constancy test.
See 4-2. “How to Customize the Constancy Test Setting”.
Input Lamb Set up a sensor for the monitor luminance value and
measurement. The monitor luminance value must be input
0.00-9.99.
See 5-1.
Advanced Mode User Mode